Skip to main navigation Skip to search Skip to main content

Wide dispersion multiple collector isotope ratio mass spectrometer

Research output: Contribution to journal › Article › peer-review

7 Scopus citations

Abstract

A new design for an enhanced dispersion thermal ionization multi-collector isotope ratio mass spectrometer is proposed and the results of tests with a small-scale prototype instrument are presented. The measured performance of the small-scale prototype sized for 6Li and 7Li is compared with the predictions of an ion optical model. The ion optical model is shown to accurately predict the measured mass dispersion, mass resolution, and general ion beam profile, providing confidence that the ion optical modeling can be applied to design of a full-scale instrument. The design concept includes a single magnetic sector and an electrostatic dispersion lens to magnify the mass dispersion so that full sized discrete dynode electron multipliers can be used to simultaneously monitor each isotope. A conceptual model is presented for a thermal ionization instrument capable of simultaneous measurement of seven isotopes in the 240 Da mass range with full size discrete dynode electron multipliers.

Original languageEnglish
Pages (from-to)1-9
Number of pages9
JournalInternational Journal of Mass Spectrometry
Volume241
Issue number1
DOIs
StatePublished - Feb 15 2005

Keywords

  • Isotope ratio
  • Magnetic sector mass spectrometer
  • Mass dispersion
  • Multiple collector
  • Thermal ionization

Fingerprint

Dive into the research topics of 'Wide dispersion multiple collector isotope ratio mass spectrometer'. Together they form a unique fingerprint.

Cite this