Time-resolved, dual heterodyne phase collection transient grating spectroscopy

Cody A. Dennett, Michael P. Short

Research output: Contribution to journalArticlepeer-review

33 Scopus citations

Abstract

The application of optical heterodyne detection for transient grating spectroscopy (TGS) using a fixed, binary phase mask often relies on taking the difference between signals captured at multiple heterodyne phases. To date, this has been accomplished by manually controlling the heterodyne phase between measurements with an optical flat. In this letter, an optical configuration is presented which allows for collection of TGS measurements at two heterodyne phases concurrently through the use of two independently phase controlled interrogation paths. This arrangement allows for complete, heterodyne amplified TGS measurements to be made in a manner not constrained by a mechanical actuation time. Measurements are instead constrained only by the desired signal-to-noise ratio. A temporal resolution of between 1 and 10 s, demonstrated here on single crystal metallic samples, will allow TGS experiments to be used as an in-situ, time-resolved monitoring technique for many material processing applications.

Original languageEnglish
Article number211106
JournalApplied Physics Letters
Volume110
Issue number21
DOIs
StatePublished - May 22 2017
Externally publishedYes

Fingerprint

Dive into the research topics of 'Time-resolved, dual heterodyne phase collection transient grating spectroscopy'. Together they form a unique fingerprint.

Cite this