Abstract
Parameter studies, design calculations and initial neutronic performance measurements have been completed for a new thermal neutron beamline to be used for neutron capture therapy cell and small-animal radiobiology studies at the University of Missouri Research Reactor. The beamline features the use of single-crystal silicon and bismuth sections for neutron filtering and for reduction of incident gamma radiation. The calculated and measured thermal neutron fluxes produced at the irradiation location are 9.6×108 and 8.8×108 neutrons/cm2 s, respectively. Calculated and measured cadmium ratios (Au foils) are 217 and 132. These results indicate a well-thermalized neutron spectrum with sufficient thermal neutron flux for a variety of small animal BNCT studies.
| Original language | English |
|---|---|
| Pages (from-to) | S222-S225 |
| Journal | Applied Radiation and Isotopes |
| Volume | 67 |
| Issue number | 7-8 SUPPL. |
| DOIs | |
| State | Published - Jul 2009 |
Keywords
- Activation
- BNCT
- Neutron source
- Thermal
Fingerprint
Dive into the research topics of 'Spectral performance of a composite single-crystal filtered thermal neutron beam for BNCT research at the University of Missouri'. Together they form a unique fingerprint.Cite this