Skip to main navigation Skip to search Skip to main content

GAS PHASE MOLECULAR SIMS.

  • P. J. Todd
  • , G. S. Groenewold
  • , C. P. Leibmann

Research output: Contribution to journal › Conference article › peer-review

1 Scopus citations

Abstract

Characteristic secondary ions can be observed if volatile amines are introduced into a secondary ion source in which a glycerol matrix is under ion bombardment. The abundance of these secondary ions relative to the abundance of the protonated glycerol ion has been found to follow the same trends observed for involatile analytes dissolved in glycerol; namely, longer chain amines yield more abundant secondary ions than do shorter chain amines. Therefore, surface activity is an important factor in determining the secondary ion abundance.

Original languageEnglish
Pages (from-to)247-248
Number of pages2
JournalAnnual Conference on Mass Spectrometry and Allied Topics
StatePublished - 1984

Fingerprint

Dive into the research topics of 'GAS PHASE MOLECULAR SIMS.'. Together they form a unique fingerprint.

Cite this