Abstract
A measurement system has been designed and built for the specific application of measuring the effective thermal conductivity of a composite, nuclear-fuel compact (small cylinder) over a temperature range of 100 °C to 800 °C. Because of the composite nature of the sample as well as the need to measure samples pre- and postirradiation, measurement must be performed on the whole compact non-destructively. No existing measurement system is capable of obtaining its thermal conductivity in a non-destructive manner. The designed apparatus is an adaptation of the guardedcomparative-longitudinal heat flow technique. The system uniquely demonstrates the use of a radiative heat sink to provide cooling which greatly simplifies the design and setup of such high-temperature systems. The design was aimed to measure thermalconductivity values covering the expected range of effective thermal conductivity of the composite nuclear fuel from 10W·m -1 ·K -1 to 70W·m -1 ·K -1. Several materials having thermal conductivities covering this expected range have been measured for system validation, and results are presented. A comparison of the results has been made to data from existing literature.Additionally, an uncertainty analysis is presented finding an overall uncertainty in sample thermal conductivity to be 6%, matching well with the results of the validation samples.
| Original language | English |
|---|---|
| Pages (from-to) | 311-329 |
| Number of pages | 19 |
| Journal | International Journal of Thermophysics |
| Volume | 33 |
| Issue number | 2 |
| DOIs | |
| State | Published - Jan 31 2012 |
Keywords
- Comparative method
- High-temperature measurement
- Nuclear fuel compact
- Thermal-conductivity measurement
- Uncertainty analysis
Fingerprint
Dive into the research topics of 'Design and validation of a high-temperature comparative thermal-conductivity measurement system'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver