Abstract
We investigated the feasibility of long-drift-time CdZnTe (CZT) gamma-ray detectors, fabricated from CZT material produced by Redlen Technologies. CZT crystals with cross-section areas of 5 × 5 mm2 and 6 × 6 mm2 and thicknesses of 20-, 30-, 40-, and 50-mm were configured as 3D position-sensitive drift detectors and were read out using a front-end ASIC. By correcting the electron charge losses caused by defects in the crystals, we demonstrated high performance for relatively thick detectors fabricated from unselected CZT material.
| Original language | English |
|---|---|
| Article number | 093504 |
| Journal | Applied Physics Letters |
| Volume | 108 |
| Issue number | 9 |
| DOIs | |
| State | Published - Feb 29 2016 |
| Externally published | Yes |
Fingerprint
Dive into the research topics of 'CdZnTe position-sensitive drift detectors with thicknesses up to 5 cm'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver