Skip to main navigation Skip to search Skip to main content

CdZnTe position-sensitive drift detectors with thicknesses up to 5 cm

  • A. E. Bolotnikov
  • , G. S. Camarda
  • , E. Chen
  • , S. Cheng
  • , Y. Cui
  • , R. Gul
  • , R. Gallagher
  • , V. Dedic
  • , G. De Geronimo
  • , L. Ocampo Giraldo
  • , J. Fried
  • , A. Hossain
  • , J. M. Mackenzie
  • , P. Sellin
  • , S. Taherion
  • , E. Vernon
  • , G. Yang
  • , U. El-Hanany
  • , R. B. James

Research output: Contribution to journal › Article › peer-review

30 Scopus citations

Abstract

We investigated the feasibility of long-drift-time CdZnTe (CZT) gamma-ray detectors, fabricated from CZT material produced by Redlen Technologies. CZT crystals with cross-section areas of 5 × 5 mm2 and 6 × 6 mm2 and thicknesses of 20-, 30-, 40-, and 50-mm were configured as 3D position-sensitive drift detectors and were read out using a front-end ASIC. By correcting the electron charge losses caused by defects in the crystals, we demonstrated high performance for relatively thick detectors fabricated from unselected CZT material.

Original languageEnglish
Article number093504
JournalApplied Physics Letters
Volume108
Issue number9
DOIs
StatePublished - Feb 29 2016
Externally publishedYes

Fingerprint

Dive into the research topics of 'CdZnTe position-sensitive drift detectors with thicknesses up to 5 cm'. Together they form a unique fingerprint.

Cite this