Skip to main navigation Skip to search Skip to main content

Benchtop isolation of radioactive and stable isotopes by ICP-MS

Research output: Contribution to journal › Article › peer-review

2 Scopus citations

Abstract

Our previous research has demonstrated the production of ultra-pure isotopes on the ng-μg scale with high purity >99.999% enabled by modifications to an inductively coupled plasma mass spectrometer (ICP-MS). This technique can also be used to improve radiometric measurements for alpha and beta counting using the isotopic selectivity, low-energy ion deposition technique. The results confirm the utility of this method for developing isotope tracers that can improve the precision and accuracy in trace mass spectrometry measurements. In order to increase throughput, a modified implant mechanism has been developed that can rotate thereby allowing several sequential implants without venting the back-end of the instrument. The details of this modification are discussed and results of isolating and implanting several Nd isotopes are presented.

Original languageEnglish
Pages (from-to)482-484
Number of pages3
JournalNuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms
Volume463
DOIs
StatePublished - Jan 15 2020

Keywords

  • ICP-MS
  • Ion deposition
  • Isotope dilution analysis
  • Isotope enrichment

Fingerprint

Dive into the research topics of 'Benchtop isolation of radioactive and stable isotopes by ICP-MS'. Together they form a unique fingerprint.